Konica Minolta CM-2500d Drums User Manual


 
E-79
Measuring the Average
2
Place the specimen measuring port to the specimen to be measured, and then
press C (MEAS. button).
If 2 is set for “MANUAL AVG.
TIMES” or if 3 or a higher value is
set and the standard deviation is not
below the specified one, place the
specimen measuring port to the next
place and then press
C (MEAS.
button).
The number of completed measurements including the one cur-
rently in progress and standard deviation will be displayed during
measurement.
When the specified number of meas-
urements has been completed or the
standard deviation drops below the
specified one, the averaging result
will be displayed on the LCD.
M
emo
To exit manual averaging and display the results before the specified number of measurements is completed or
before the specified standard deviation is reached, press A.
• If does not appear due to the auto power save function, this may be due the fact that it can take time
to start measurement after C is pressed.