Enclosure Double molded, waterproof (IP67)
Shock (Drop Test) 2 meters (6.5 feet)
Diode Test Test current of 0.9mA maximum, open circuit voltage 2.8V DC typical
Continuity Check Audible signal will sound if the resistance is less than 100 (approx.), test
current <0.35mA
Temperature Sensor Requires type K thermocouple
Input Impedance >10M VDC & >10M VAC
AC Response True rms
ACV Bandwidth 40Hz to 1kHz
Crest Factor 3 at full scale up to 500V, decreasing linearly to 1.5 at 1000V
Display 6,000 count backlit liquid crystal with bargraph
Overrange indicationOL” is displayed
Auto Power Off 15 minutes (approximately) with disable feature
Polarity Automatic (no indication for positive); Minus (-) sign for negative
Measurement Rate 2 times per second, nominal
Low Battery Indication “ ” is displayed if battery voltage drops below operating voltage
Battery One 9 volt (NEDA 1604) battery
Fuses mA, µA ranges; 0.8A/1000V ceramic fast blow
A range; 10A/1000V ceramic fast blow
Operating Temperature 5C to 40C (41F to 104F)
Storage Temperature -20C to 60C (-4F to 140F)
Operating Humidity Max 80% up to 31C (87F) decreasing linearly to 50% at 40C (104F)
Storage Humidity <80%
Operating Altitude 2000 meters (7000ft) maximum.
Weight 342g (0.753lb) (includes holster).
Size 187 x 81 x 50mm (7.36” x 3.2” x 2.0”) (includes holster)
Safety This meter is intended for origin of installation use and protected, against
the users, by double insulation per EN61010-1 and IEC61010-1 2
nd
Edition
(2001) to Category IV 600V and Category III 1000V; Pollution Degree 2.
The meter also meets UL 61010-1, 2
nd
Edition (2004), CAN/CSA C22.2 No.
61010-1 2
nd
Edition (2004), and UL 61010B-2-031, 1
st
Edition (2003)
Approvals UL CE
UL Listed The UL mark does not indicate that this product has been evaluated for the
accuracy of its readings.