
Columns and Fittings
Inlet inserts
27
Note: For the liner and insert for an HP Series 530 capillary column, if
the column is already installed, a new insert may not seat properly in the
liner; the column may prevent it from dropping completely into the liner.
If the insert does not drop completely into the liner, do not force it
(either the liner or the column may shatter); instead, remove the column,
seat the insert, and then replace the column.
5. Replace the septum and septum retainer nut.
Split/splitless or split-only capillary inlet inserts
Viton O-ring (Preferred)
Viton O-ring
Split Use Splitless Use
Split/Splitless and Split-Only Capillary Inlet and Inserts
Figure 1-10
A specific inlet insert is required, depending upon the particular
sampling mode. Specific sampling modes include:
Split, for major•componentanalyses
Purged splitless, for trace•componentanalyses