ACCESSORIES
Electric Portable Patient Lift 46 Part No. 1145810
NOTE:Stablebeingdefinedastheweightfluctuatingtwotenthsofapound.Forexample,a
patientweighingonehundredpounds,thescalewillfluctuatebetween99.8and100.2untilthe
LOCKkeyispressed.Fluctuationoftheweightdisplayedisnormalasnotedabove.Pressthe
LOCKbuttontolocktheweight.
NOTE:TheUNITSbuttoncanbepressedtotogglebetweenunitsofpoundsandkilograms.This
isindicatedbylborkgappearinginthedisplaywindow.
8. Theliftmaynowbeloweredandtheslingremovedfromthepatient.
NOTE:Thepatientʹsweightwillcontinuetobeseeninthedisplaywindow.Thedisplaywillturn
offautomaticallyafteratwominuteperiodofnon‐use[nochangesinweightexceedingfivepounds
(twokilograms)].YoucanNOTadjustthetimedelayforautomaticshutoff.Afterthedisplayhas
turnedoff,theweightmayberecalledbypressingtheON/OFFbutton.Theunitcanbeturnedoff
bypressingtheON/OFFbuttonasecondtime.
Replacing the Battery
NOTE:Forthisprocedure,refertoFIGURE 8.4.
NOTE:Thescaleispoweredbyaninevoltalkalinebatterythatshouldprovideapproximately1500
readingsbeforeneedingreplacement.
Whenbatteryreplacementisneeded,LOBATwillappearonthedisplay.Performthe
following:
1. Slidethebatterydooropeninthedirectionofthearrow.
2. Removeexistingbattery.
3. Installthenewbattery.
4. Reinstallthebatterydoor.
FIGURE 8.4 Replacing the Battery
Calibrating the Reliant Scale
NOTE:TheReliantScalewillbepre‐calibratedatthefactorywiththeloadcell.Shoulditbe
necessarytore‐calibratethescale,followtheinstructionsoutlinedbelow.
1. Thepatientandtheslingmustberemovedfromthescaletoproperlycalibratethe
ReliantScale.Forremovingthepatientinstructions
refertoOperatingthePatientLift
onpage 22.ForslingdetachmentinstructionsrefertoLiftingthePatient
onpage 26.
Battery
Door
9V Battery
Assembled
Scale
Arrow