![](https://pdfstore-manualsonline.prod.a.ki/pdfasset/0/5f/05f671ca-ea05-4aac-9ee1-4f864f41e3db/05f671ca-ea05-4aac-9ee1-4f864f41e3db-bg1d.png)
EN-23
CD-11 Indicator
GEOGRAPHICAL FACTOR (For Europe Only)
Press the G/N/T/
Menu
button, gEO 19 is displayed. This is the current geo-
graphical adjustment value.
The geo factor includes settings from 0 to 31 and is used to compensate for
varying gravity at different geographical areas (complete geographical adjust-
ment settings are listed in the following table).
NOTE:
Only an authorized manufacturer’s representative or certified verification
personnel may make these changes.
Changing the geographical setting alters the calibration values.
Press the Print/
Units
button until the desired geographical adjustment value is
reached. The factory default setting is gEO 19
SAVE
Press the G/N/T/
Menu
button to end this block, SAVE is displayed.
Press the G/N/T/Menu button to save the geographical factor setting. The next
menu SEt uP is displayed.
or
Press the Print/
Units
button to go back to the CAL menu without saving
changes to the geographical setting.
QUIT
To exit from the CAL menu, press the Print/
Units
button to advance to Quit.
Then press the G/N/T/
Menu
button to go back to the weighing mode.