
Maintaining, Testing, and Troubleshooting Your HeartStart FR2+
4-15
Philips Medical Systems
4
Status Indicator:
SOLID RED X
Audio signal:
NONE
• The battery is missing or completely
depleted.
• The training & administration pack is
being used in the administration
function (the solid red X is normal in
this case) or has been left in the FR2+
by mistake.
• A self-test detected a failure.
• Install a new M3863A FR2
standard battery or a charged
M3848A FR2+ rechargeable
battery in the FR2+ and perform
the battery insertion test (BIT).
• Remove the training &
administration pack and install a
battery.
• Remove and reinstall the battery
and perform the battery insertion
self-test. If it fails, install a new
M3863A FR2 standard battery or
a charged M3848A FR2+
rechargeable battery and repeat
the test. If it fails again, do not use
the FR2+.
Status Indicator:
SOLID RED X
Audio signal:
CHIRPING
• The training & administration
pack is being used in the
ADMINISTRATION function
and more than 10 minutes have
passed without user interaction
(button press or pads change).
• The training & administration pack is
being used in the
TRAINING function
and more than 30 minutes have
passed without user interaction
(button press or pads change).
• To continue using the training &
administration pack, press any
button (except On/Off).
• To return the FR2+ to standby
mode, remove the Pack and install
a battery.
Status Indicator:
NONE
The FR2+ has been physically damaged. • Check for visible damage. Do not
use the FR2+ if it appears to be
damaged.
• Remove and reinstall the battery to
perform the battery insertion
self-test. If it fails, install a new
M3863A FR2 standard battery or
a charged M3848A FR2+
rechargeable battery and repeat
the test. If it fails again, do not use
the FR2+.
symptom possible cause recommended action