
F-2
Philips Medical Systems
HEARTSTART M5066A
Patient treatment protocol options
parameter settings default default description
“call EMS” voice
reminder timing
• At power on (when
the user turns on
the HeartStart)
• At power on and at
the start of the first
patient care pause
• At the start of the
first patient care
pause
•No reminder
At the start of
the first
patient care
pause
Provides a voice reminder to make
sure emergency medical services have
been called, at the start of the first
patient care pause.
shock series 1, 2, 3, 4 1 The automatic protocol pause for CPR
is activated each time a shock is
delivered.
*
During the protocol pause, the
HeartStart does not perform rhythm
analysis.
The length of the protocol pause after
a shock series is completed is
determined by the protocol pause
timer setting.
shock series interval
(minutes)
1.0, 2.0,
∞ (infinity)
1.0 A delivered shock must occur within 1
minute of the previous shock to be
counted as part of the current shock
series.
NOTE: This parameter is only applicable
when the shock series is not configured to
the default 1 shock.
* A shock series begins when a shock is delivered after the HeartStart is turned on. A new shock series begins after a
protocol pause. If shock series is configured for 2 or more, a new shock series also begins if the time since the previous
shock exceeds the shock series interval setting.