6.0 USER PARAMETERS
6.1 AUTO POWER OFF
Press [Func] and [Tare/Zero] keys at the same time and display will
show “ t x”.
t (time)
x (digit 0-9, total 10 settings)
Press the [Func] key. x will change from 0 to 9.
X=0……………no automatically power off function
X=1……………auto power off after 2 minutes
X=2……………auto power off after 4 minutes
……………………………………………….
X=9……………auto power off after 18 minutes
(X is increased by 2 minutes with each change)
When the desired time is displayed, press [Hold] key to confirm. If you
wish to set the next parameter, press the [Hold] key and enter the Zero
setting. If you don’t wish to continue with the setting, press the [Func]
key to return to the weighing function.
6.2 CALIBRATION
6.2.1 ZERO SETTING
Press the [Hold] key to confirm and enter the zero setting mode.
When “000000” is displayed, press the [Hold] key to confirm.