SetupMode
EnteringSetupMode
WiththemeterswitchedOFF,pressandholdthePOWERSETbuttonfortwo(2)secondsto
enterSetupmode.Five(5)parametersareavailable:
P10:DataloggerSamplingRate
P20:99ReadingMemoryErase
P30:UnitsSelection
P40:RealTimeClock
P50:HumidityOffset
Usetheup/downbuttonstostepthrough
theselections.
P10:DataloggingSamplingRate
1. Onceinthesetupmode,attheP10RATEdisplay,momentarilypresstheENTERbutton
toaccesstheP10menu.
2. TheHOURSdigitsshouldnowbeflashing.Usethearrowbuttonstoselectthedesired
hourssetting.PresstheENTERbuttontoconfirmtheentry.
3. TheMINUTESdigitsshouldnowbeflashing.Usethearrowbuttonstoselectthedesired
setting.PresstheENTERbuttontoconfirmtheentry.
4. TheSECONDSdigitsshouldnowbeflashing.Usethearrowbuttonstoselectthedesired
setting.PresstheENTERbuttontoconfirm
theentry.
5. ThedisplayshouldnowreturntothemainP10RATEdisplay.
6. UsethearrowkeystosteptoanotherparameterorpressESCtoexittheSETUPmode.
P20:CLR(99ReadingMemoryClear)
1. Onceinthesetupmode,attheP20CLRdisplay,momentarily
presstheENTERbuttonto
accesstheP20menu.
2. ThewordNOorYESwillbeflashing.UsethearrowkeystoselectNO(donoterasedata)
orYES(erasedata).WhenthedesiredcommandisshownpresstheENTERbuttonto
executethecommand.IfYESwasselected,
theentire99Readingmemorybankwillbe
erased.IfNOwasselected,thedatawillnotbeerasedandwillremaininmemory.
3. ThedisplayshouldreturntothemainP20CLRdisplayoncethecommandisexecuted.
4. Usethearrowkeystosteptoanotherparameter
orpressESCtoexittheSETUPmode.
P30:UNIT
1. Onceinthesetupmode,attheP30UNITdisplay,momentarilypresstheENTERbuttonto
accesstheP30menu.
2. Usethearrowbuttonstoselecttheimperialormetricunitset.Theavailableunitsare:air
velocity(m/s,
fpm),temperature(C,F),airvolume(CMM,CFM),andareasize(cm²,
inch²).
3. PressENTERtoconfirmtheselection.
4. UsethearrowkeystosteptoanotherparameterorpressESCtoexittheSETUPmode.