Getting Started
80E00 Electrical Sampling Modules User Manual
9
After installing a sampling module or after moving a sampling module from one
compartment to another, you should run compensation from the Utilities menu to
ensure the instrument meets it specifications. Also run a compensation (accessed
from the Utilities menu) when doing the following:
H Installing an 80E00 sampling-module extender between the instrument and
an 80E00 sampling module, where none was used before.
H Removing an 80E00 sampling-module extender between the instrument and
an 80E00 sampling module, where one had been used before.
H Exchanging an extender for one of a different length.
For instructions on running a compensation, see Optimizing Measurement
Accuracy in the Online Help for your main instrument.
Compensation