Specifications
80E00 Electrical Sampling Modules User Manual
49
Table 7: Electrical sampling modules - Signal acquisition (cont.)
Specifications Characteristics
n Rise tim e
4
Sampling module Rise time
80E01 ≤ 7 ps, typical
80E02 ≤ 28 ps
80E03 and 80E04 ≤ 17.5 ps
80E06 ≤ 5.0 ps, typical
n Analog bandwidth
5
Sampling module Bandwidth
80E01 50 GHz
80E02 12.5 GHz, typical
80E03 and 80E04 20 GHz, typical
80E06 65 GHz
70 GHz, typical
Step response
7
Sampling module Aberrations, step transition
6
p
p
aberrations
7
,typical
80E02, 80E03 and 80E04
±3% or less over the zone 10 ns
to 20 ps before step transition
+10%, --5% or less for the first
300 ps following step transition
±3% or less over the zone
300 ps to 5 ns following step
transition
±1% or less over the zone 5 ns
to 100 ns following step transition
±0.5% after 100 ns following
step t ransi tion
80E01
±3% or less over the zone 10 ns
to 20 ps before step transition
+12%, --5% or less for the first
300 ps following step transition
+5.5%, --3% or less over the zone
300 ps to 3 ns following step
transition
±1% or less over the zone 3 ns
to 100 ns following step transition
±0.5% after 100 ns following
step t ransi tion
Step response
overshoot
7
,typical
80E06 + 5% or less for the first 300 ps
following step transi tion