Agilent Technologies 60501B Drums User Manual


 
8
Table 60501-4. Calibration Information
Ranges and
Calibration Points
Variables Variables
Value
Power Supply
Settings
Current
Shunt
High Current Range Hi_curr_rng 30 5 V/31 A 100 A
High Current Offset Hi_curr_offset 0.013
Low Current Range Lo_curr_rng 3 5 V/10 A 15 A
Low Current Offset Lo_curr_offset 0.013
Voltage Range N/A N/A 61 V/2 A N/A
Voltage Hi point Volt_hipt 60
Voltage Lo point Volt_lopt 2.7
Low Resistance Range Lo_res_rng 2 15 V/5.5 A 15 A
Low Resistance Hi point Lo_res_hipt 1.9
Low Resistance Lo point Lo_res_lopt 0.067
Middle Resistance Range Mid_res_rng 20 10.9 V/8 A 15 A
Middle Resistance Hi point Mid_res_hipt 60
Middle Resistance Lo point Mid_res_lopt 2.1
High Resistance Range Hi_res_rng 2002 60 V/5 A 15 A
High Resistance Hi point Hi_res_hipt 200
High Resistance Lo point Hi_res_lopt 24