Agilent Technologies 60501B Drums User Manual


 
4
Table 60501-1 Specifications and Supplemental Characteristics (continued)
External Analog Programming 0 to 10 V (dc or ac):
Bandwidth: 10 kHz (3 db frequency)
Accuracy: ± 4.5% ± 40 mA (0 to 3 A range)
± 4.5% ± 130 mA (0 to 30 A range)
± 0.8% ± 200 mV (0 to 60 V range)
Temperature Coefficient: 100 ppm/
°
C ± 3 mA/
°
C (current ranges)
100 ppm/
°
C ± 1 mV/
°
C (voltage range)
External Current Monitor (0 to 10 V):
Accuracy: ± 4% ± 40 mA (referenced to analog common)
Temperature Coefficient: 60 ppm/
°
C ± 3 mA/
°
C
External Voltage Monitor (0 to 10 V):
Accuracy: ± 0.25% ± 40 mV (referenced to analog common)
Temperature Coefficient: 50 ppm/
°
C ± 0.2 mV/
°
C
Remote Sensing: 5 Vdc maximum between sense and input binding posts
Maximum Input Levels:
Current: 30.6 A (programmable to lower limits)
Voltage: 75 V
Minimum Operating Voltage: 2 V (derated to 0 V at 0 A)
PARD (20 Hz to 10 MHz noise):
Current: 2 mA rms/20 mA p-p
Voltage: 5 mV rms
DC Isolation Voltage: ± 240 Vdc between + or - input binding post and chassis ground
Digital Inputs:
Vlo: 0.9 V maximum at Ilo = -1 mA
Vhi 3.15 V minimum (pull-up resistor on input)
Digital Outputs:
Vlo: 0.72 V maximum at Ilo = 1 mA
Vhi: 4.4 V minimum at Ilo - 20 µA
SUPPLEMENTAL CHARACTERISTICS
Programmable Slew Rate (For any given input transition, the time required will be either the total slew time or a minimum
transition time, whichever is longer. The minimum transition time increases when operating with input currents under 1 A.
The following are typical values; ±25% tolerance):