Agilent Technologies 60501B Drums User Manual


 
2
Table 60501-1. Specification and Supplemental Characteristics
SPECIFICATIONS
DC Input Rating:
Current: 0 to 30 A
Voltage: 3 to 60 V (minimum dc operation from 0 to 2 V for 0 to 30 A)
Power: 150 W at 40
°
C (derated to 112 W at 55
°
C)
A. OPERATING CHARACTERISTICS B. DERATED CURRENT DETAIL
Constant Current Mode:
Ranges: 0 to 3 A; and 0 to 30 A
Accuracy: (after 30 second wait):
±
0.1% ± 40 mA (both ranges)
Resolution: 0.8 mA (3 A range); 8 mA (30 A range)
Regulation: 10 mA (both ranges)
Temperature Coefficient: 100 ppm/
°
C ± 3 mA/
°
C (both ranges)
Constant Resistance Mode:
Ranges: 0.067 to 2
; 2
to 2 k
; and 20
to 10 k
Accuracy: ± 0.8% ± 16 m
with
6 A at input (2
range);
± 0.3% ± 5 mS with
6 V at input (2 k and 10 k
ranges)
Resolution: 0.54 m
(2
range); 0.14 mS (2 k
range); 0.014 mS (10 k
range)
Regulation: 10 mV with remote sensing (2
range); 10 mA (2 k and 10 k
ranges)
Temperature Coefficient: 800 ppm/
°
C ± 0.8 m
/
°
C (2
range);
300ppm/°C ± 0.5 mS/°C (2 k and 10 k ranges)
Constant Voltage Mode:
Range: 0 to 60 V
Accuracy: ± 0.1% ± 50 mV
Resolution: 16 mV
Regulation: 5 mV (remote sense); 40 mV (local sense)
Temperature Coefficient: 100 ppm/°C ± 5 mV/°C