Emerson SPS 4001B Recording Equipment User Manual


 
Instruction Manual
IM-106-340AC-1, Rev. 1.0
April 2005
SPS 4001B
3-4
SEMI-AUTOMATIC
CALIBRATION
Semi-automatic calibrations are operator initiated and can be performed using
the Oxymitter 4000 keypad, HART handheld communicator/AMS software, or
a remote contact. In addition, the calibration gases must be permanently
piped to the Oxymitter 4000.
Oxymitter 4000 Keypad
A semi-automatic calibration can be initiated by pressing the CAL button on
the Oxymitter 4000 keypad. For more information, refer to the Oxymitter 4000
Oxygen Transmitter Instruction Manual.
HART Handheld Communicator/AMS Software
A semi-automatic calibration can be initiated by connecting the HART
handheld communicator, or AMS software, to the Oxymitter 4000 4-20 mA
signal line and using the HART communicator keypad or computer keyboard
to access the applicable calibration menu. Refer to the Oxymitter 4000
Oxygen Transmitter Instruction Manual or HART documentation for more
information.
Remote Contact
A semi-automatic calibration can be initiated using a remote contact such as a
customer's control system. The remote contact processes the calibration
command on a PC and sends the signal to the Oxymitter 4000. For more
information on remote-site calibrations, refer to the documentation for the
system in use.