HP (Hewlett-Packard) EPM E4419A Music Mixer User Manual


 
HP4402.book : chapter2.fm 17 Wed Apr 23 15:29:05 1997
Power Meter Operation
Making Measurements using Sensor Calibration Tables
HP EPM-442A User’s Guide 2-17
The power meter is shipped with a set of predefined sensor calibration
tables. The data in these tables is based on statistical averages for a range
of Hewlett-Packard power sensors.
These power sensors are:
DEFAULT
1
HP 8481A
HP 8482A
2
HP 8483A
HP 8481D
HP 8485A
R8486A
Q8486A
R8486D
HP 8487A
There are also ten sensor calibration tables named CUSTOM_0 through
CUSTOM_9 which do not contain any data when the power meter is
shipped from the factory.
You cannot delete any of the 20 sensor calibration tables. Each sensor
calibration table can contain a maximum of 80 frequency points each. If
you want a new sensor calibration table you must edit an existing one.
To edit a table:
1. Select one using the and hardkeys to scroll through the
displayed tables. When the table you want to edit is highlighted
press . The “Edit Cal” screen is displayed as
shown in Figure 2-2 on page 2-16. In this screen you can modify,
add and remove frequency and calibration factors and also edit the
table name. Use the , , and hardkeys to move
between the table name and its frequency and calibration factors.
A frequency in the range of 0.1 MHz to 999.999 GHz can be
entered.
1. DEFAULT is a sensor calibration table in which the reference calibration
factor and calibration factors are 100%. This sensor calibration table can be
used during the performance testing of the power meter.
2. The HP 8482B and HP 8482H power sensors use the same data as the
HP 8482A.
Edit Cal Table