7532 Mounting
3.4 Installation instructions
3.4.1 Mounting kit
In addition to the tool needed for flange mounting, you will require the following tool:
4 mm Allen wrench for turning the housing.
3.4.2 Installation in stilling well
Optimum mounting position for 7532 RTG
Recommendations for the stilling well
Metal (no enamel coating, plastic on request).
•Constant diameter.
Weld seam as smooth as possible and on the same axis as the slots.
Slots offset 180° (not 90°).
Slot width respectively diameter of holes max. 1/10 of pipe diameter, de-burred.
Length and number do not have any influence on the measurement.
Select antenna extension as big as possible. For intermediate sizes (e.g. 180 mm)
select next larger antenna extension and adapt it mechanically. Maximum gap allowed
between the antenna/horn of 7532 RTG and the inside of the stilling well is 5 mm (3/
16").
The antennna extension of the 7532 RTG is mounted with defined pressure.
It is strongly recommended not to dismantle this antenna.
Dimensions of a nozzle for manual gauging must be adapted to the dimensions of
the horn antenna used, compare to page page 23.
Stilling wells with pipe expansion
When using a 7532 RTG, a gradual pipe expansion from DN150 to DN200 / DN200
to DN250 / DN250 to DN300 is acceptable. In such cases, the top end of the piping
must have a minimum length of 0.5 m in front of the expansion (see table on page 18).
Ideally, a dip socket is used for this.
Greater expansion of the piping (e.g. DN150 to DN300) is possible if the top end of
the pipe is extended accordingly (see table on page 18).
Avoid pipe expansions at right angles.
Standard installation
When mounting in the stilling well, please
observe engineering tips on page 14 and
the following points:
After mounting, the housing can be
turned 350° in order to simplify access to
the display and the terminal
compartment.
planar axis vertical to flange.
Measurements can be performed
through an open ball valve without any
problems.
L00-FMR532xx-17-00-00-en-003