Inserting Oscilloscope Waveforms
Command
:MACHine{1|2}:TWAVeform:INSert <module_spec>,
<label_name>
This inserts a waveform from an oscilloscope to the timing waveforms display.
<module_spec> {1|2|3|4|5|6|7|8|9|10} slot in which the oscilloscope master card is
installed
<label_name> string of one alpha and one numeric character
Example OUTPUT XXX;":MACHINE1:TWAVEFORM:INSERT 3, ’C1’"
MLENgth
Command :MACHine{1|2}:TWAVeform:MLENgth <memory_length>
The MLENgth command allows you to specify the analyzer memory depth.
Valid memory depths range from 4096 samples through the maximum system
memory depth minus 8192 samples (HP 16554A/HP 16555A) or the
maximum system memory depth minus 16384 (HP 16555D). Memory depth
is affected by acquisition mode. If the
<memory_depth> value sent with the
command is not a legal value, the closest legal setting will be used.
<memory_length> {4096 | 8192 | 16384 | 32768 | 65536 | 131072 | 262144
| 516096 | 524288 | 1040384 (HP 16554A)
| 524288 | 1040384 | 1048576 | 2088960 (HP 16555A)
| 524288 | 1048576 | 2080768 | 2088960 | 4177920
(HP 16555D)}
Example OUTPUT XXX;":MACHINE1:TWAVEFORM:MLENGTH 262144"
TWAVeform Subsystem
MLENgth