ARMLine
Command :ARMLine MACHine<N>
The ARMLine command selects which machine generates the arm out signal
on the IMB (intermodule bus). This command is only valid when two
analyzers are on. However, the query is always valid.
<N> {1|2}
Example OUTPUT XXX;":ARMLINE MACHINE1"
Query :ARMLine?
If the analyzer is set up for OR’d triggering, then the ARMLine query returns
an empty string. This cannot be used for setting up OR’d triggering.
Returned Format [:ARMLine]{MACHine<N>|}<NL>
Example OUTPUT XXX;":ARMLine?"
DBLock
Command :DBLock {PACKed | UNPacked}
The DBLock command specifies the data block format that is contained in
the response from a :SYSTem:DATA? query. See Chapter 17 for more
information on the :SYSTem:DATA command and query.
The PACKed option (default) uploads data in a compressed format. This
option is used to upload data for archiving, or for reloading back into the
analyzer. When an analyzer configuration is saved to disk, the PACKed data
format is always used (regardless of the current DBLock selection).
Module Level Commands
ARMLine