A SERVICE OF

logo

Additionally, the following sections may also be included, depending on what
is available:
"SYMBOLS A "
"SYMBOLS B "
"INVASM A "
"INVASM B "
<block data> <block length specifier><section>...
<block length
specifier
#8<length>
<length> the total length of all sections in byte format (must be represented with 8
digits)
<section> <section header><section data>
<section
header>
16 bytes in the following format:
10 bytes for the section name
1 byte reserved
1 byte for the module ID code (34 for the HP 16554A and HP 16555A/D
logic analyzer)
4 bytes for the length of the section data in bytes
<section data> format depends on the type of data. The total length of a section is 16 (for
the section header) plus the length of the section data. So when calculating
the value for
<length>, don’t forget to include the length of the section
headers. The format of the setup block is not affected by the DBLock
command setting.
Example OUTPUT XXX;"SETUP" <block data>
Query :SYStem:SETup?
The SYStem:SETup query returns a block of data that contains the current
configuration to the controller.
Returned Format [:SYStem:SETup] <block data><NL>
DATA and SETup Commands
SYSTem:SETup
17–14