Using the Mixers with the 856X Series Spectrum Analyzers
Operation
Identify signals in wide frequency spans
7.
SIG ID AT MKR identifies signals in wide frequency spans, using harmonic search. SIG ID
AT MKR
automatically determines the proper frequency of a signal and displays its
value on the spectrum analyzer.
Activating
SIG ID AT MKR on an image of the signal will yield a reading in the active
block, as shown in Figure 2-14. The true signal frequency is given with an identifier
that the marker resides on an image. If the marker is placed on a true signal, and
SIG ID AT MKR is activated, the signal frequency will appear without the IMAGE
notation, as shown in Figure 2-15.
8. To activate
SIG ID AT MKR:
a. Place a marker on a signal.
b. Press
AUX CTRL, EXTERNAL MIXER, SIGNAL IDENT, SIG ID AT MKR.
9. To exit the external mixer mode, press
AUX CTRL, then INTERNAL MIXER.