Agilent Technologies 11970-90031 Music Mixer User Manual


 
52 Chapter 3
Performance Tests
Performance Test Procedures
Performance Tests
Table 3-2. Recommended Test Equipment for 11970A
Instrument Critical Specifications Recommended
Model
Spectrum Analyzer LO and IF ranges compatible with mixer 8566B
Synthesized Sweeper Frequency: 8 to 13.5 GHz
Output Level: >0 dBm
8340A
Amplifier Output Level: >18 dBm leveled
Frequency Range: 3 to 5 GHz
11975A
Amplifier Frequency Range: 8 to 13.5 GHz
Output Level >15 dBm
8349A
Power Meter Compatible with Power Sensor 436A
Power Sensor SWR: <2.0 R8486A
Power Sensor SWR: <1.3 @ 6 GHz 8485A
Directional Coupler
1
1. Calibration data for the coupling ration between the output arm and the auxil-
iary arm is necessary for accurate measurements.
Coupling: 20 dB
Directivity: >40 dB
Primary Arm SWR: <1.05
Auxiliary Arm SWR: <1.2
R752D
Isolator Insertion Loss: <1.5 dB
Isolation: >20 dB
SWR: <1.2
P/N 0960-0082
Frequency Tripler Input Power: >10 dBm minimum
Conversion Loss: <15 dB
Spacekom Microwave
2
TKa-1
2. Honeywell, Inc. Spacekom Microwave Center, Santa Barbara, CA
Adapter SMA female to SMA female P/N 1250-1158
Adapter SMA female to Type N female P/N 1250-1404
Adapter (2 required) Type N male to SMA female P/N 1250-1250
Cables (3 required) Connectors: SMA male P/N 5061-5458
Cable Connectors: SMA male
Loss: <1.0 dB @ 20 GHz
P/N 8120-4396
Cable Connectors: BNC male 11086A