Real-Ear Measurements 123
5. Press [<, >] to select INS. EAR.
6. Press [EXIT] to return to Audiogram Entry Screen.
7. Press [F2] repeatedly until RECD (or REDD) is selected. You will see an addi-
tional column of average RECD information for the selected age appear on
the screen.
8. Make sure that SHOW RECD is displayed above [F5]. If it isn’t, press [F5] to
toggle the display.
9. Look at the bottom right corner of the LCD screen. If you see the message
“Earphone not calibrated,” you will need to follow the instructions found in
Appendix C for calibrating the insert earphones. See Figure 5.3.7.1.
10. Press [F3] to select MEASURE RECD.
11. Plug the insert earphone into the “earphone” jack in the back of the FP35.
Figure 5.3.6.1—This screen indicates that the coupler part of the RECD needs to be taken.
5.3.6.2 Client Setup for RECD
This section describes how to set up the client for the real-ear portion of the
RECD measurement.
1. Insert the probe microphone into your client’s ear. See Figure 5.3.6.2A.
2. Insert custom earmold or foam eartip coupled to the insert earphone into
your client’s ear. See Figure 5.3.6.2B.