A SERVICE OF

logo

26 FONIX FP35 Hearing Aid Analyzer
2.3.1 Customizing Your Instrument Default Settings
All of the settings available in the Main, Advanced, and Resets Default Settings
menus can be saved to the FP35 analyzer’s power-on default condition. In
recognition that the FP35 might be used by several different users who might
prefer different default settings, three different sets of defaults can be stored to
SETTING 1, SETTING 2, and SETTING 3, respectively.
To store the default settings:
1. Press [MENU] from the Opening screen to open the Default Settings menu.
2. Use [F2] to select the desired user number.
3. Set up the Default Settings menu as you like it. See Sections 2.3.5, 2.3.6, and 2.3.7
for an explanation of the different available settings.
4. Press [F5] to store the menu.
5. Press [START] to complete the action and store the new defaults. Pressing
any other key will cancel the action.
To change the setting number:
Use [F2] in the Default Settings menu to change the setting number. This
switches to the default settings programmed with that setting number. To save
new default settings to this setting number, follow the instructions outlined
above.
To quickly switch between the default setting programs without entering the
Default Settings menu, you can change the setting number in the Opening
screen using the [F1] key.
To reset default settings:
• Press[F1]toreturnallthesettingsintheDefaultSettingsmenutotheir
factory default condition.
• Press[F4]toreturnallthesettingstothedefaultconditionspecifiedfor
the selected setting number.
2.3.2 Switching between Easy and Advanced User Levels
The FP35 analyzer has “Easy” and “Advanced” user levels. When in Easy Mode,
the measurement screens have a cleaner look with less information about the
test conditions. The more advanced menu items have also been eliminated in
Easy Mode.
To switch between user modes:
1. Press [MENU] in the Opening screen to open the Default Settings menu.
2. Press [NEXT] to go to the Advanced Default Settings menu.
3. Use the [∨, ∧] keys to select USER LEVEL
4. Use [<, >] to choose either EASY or ADVANCED.
5. Press [F5] and [START] to save the new setting as default.
6. Press [EXIT] to close the Default Settings menu and return to the Opening screen.