174 FONIX FP35 Hearing Aid Analyzer
To view the target in numerical format:
1. Set the USER LEVEL to ADVANCED in the Default Settings menu. See
Section 2.3.2 for details.
2. Press [MENU] from the Coupler Target or Coupler EarSim screen to enter the
main menu.
3. Press [NEXT] to enter the Advanced menu.
4. Use [∧,∨] to select TABULATION.
5. Use [<, >] to select TARGET.
6. Use [EXIT] to return to the Target Coupler screen. The target data will now
be displayed. See Figure 6.3.2.
Figure 6.3.2—Numerical display of target data
6.3.3 Tying the Target to a Measurement Curve
By default, the target is always “tied” to the second measurement curve. That
is, the source type and amplitude of that curve are used to create the target.
Whenever you change the source amplitude or type of that curve, the target will
alter as soon as you take a measurement.
If you would like to tie the target to a different measurement curve: