A SERVICE OF

logo

SPERiod
Command :MACHine{1|2}:TTRigger:SPERiod <sample_period>
The SPERiod command allows you to set the sample period of the timing
analyzer.
<sample_period> real number from 4 ns (HP 16554A) or 2 ns (HP 16555A/D) to 8 ms
Example
OUTPUT XXX;":MACHINE1:TTRIGGER:SPERIOD 50E9"
Query :MACHine{1|2}:TTRigger:SPERiod?
The SPERiod query returns the current sample period.
Returned Format
[:MACHine{1|2}:TTRigger:SPERiod] <sample_period><NL>
Example OUTPUT XXX;":MACHINE1:TTRIGGER:SPERIOD?"
TTRigger (TTRace) Subsystem
SPERiod
12–19