HP4402.book : chapter3.fm 18 Wed Apr 23 15:29:05 1997
Menu Reference
Front Panel Menu Reference "System/Inputs" Menu
3-18 HP EPM-442A User’s Guide
The softkey menu structured to this hardkey affects the system setup and
the channel setup.
Press this hardkey to access the “System/Inputs” menu. This menu allows
you to set up a variety of conditions for your power meter, for example:
HP-IB address, sensor calibration tables, averaging, duty cycle, range,
offset, service. It also allows you to review errors.
Press this softkey to access a menu which allows you to set the
HP-IB address and select the HP-IB command set you want to
use.
Press this softkey to set the HP-IB address of your power
meter between 0 and 30. The default value is 13. A pop up
window appears on the screen. Use the , , and
hardkeys to change the value. To confirm your choice
press . Refer to “HP-IB Address”, on page 2-42 for
further information.
Press this softkey to access a menu and screen which allows you to
select and edit sensor calibration tables. Use the and
hardkeys to scroll through the displayed sensor calibration tables.
Sensor calibration tables are only required for the HP 8480 series
power sensors.
Press this softkey to access a menu and screen which allows
you to modify, add and remove frequency and calibration
factors from the currently selected table and modify the table
name. Use the , , and hardkeys to move
between the name, frequency and calibration factors. Press
, or as follows:
Press this softkey to modify the highlighted parameter,
which is either a frequency, calibration factor or table name.
A pop up window displays the parameter. Use the ,
, and hardkeys to change the parameter
value. To confirm your choice press .
System
Inputs
HP-IB
HP-IB Addr
Enter
Sensor Cal Tables
Edit Cal Table
Change Insert Delete
Change
Enter