Error Messages
Error Messages
HP EPM-442A User’s Guide 4-5
-113 Undefined header
A command was received that is not valid for this power meter. You may
have misspelled the command or it may not be valid command. If you
are using the short form of the command, remember that it may contain
up to four letters.
For example, TRIGG:SOUR IMM.
-121 Invalid character in number
An invalid character was found in the number specified for a parameter
value.
For example, SENS:AVER:COUN 128#H.
-123 Exponent too large
A numeric parameter was found whose exponent was larger than
32,000.
For example, SENS:COUN 1E34000.
-124 Too many digits
A numeric parameter was found whose mantissa contained more than
255 digits, excluding leading zeros.
-128 Numeric data not allowed
A numeric value was received within a command which does not accept
a numeric value.
For example, MEM:CLE 24.
-131 Invalid suffix
A suffix was incorrectly specified for a numeric parameter. You may
have misspelled the suffix.
For example, SENS:FREQ 200KZ.
-134 Suffix too long
A suffix used contained more than 12 characters.
For example, SENS:FREQ 2MHZZZZZZZZZZZ.
-138 Suffix not allowed
A suffix was received following a numeric parameter which does not
accept a suffix.
For example, INIT:CONT 0Hz.