HP (Hewlett-Packard) EPM-442A Music Mixer User Manual


 
HP4402.book : chapter2.fm 53 Wed Apr 23 15:29:05 1997
Power Meter Operation
Self Test
HP EPM-442A User’s Guide 2-53
HP-IB Testing
To invoke the HP-IB self test, the IEEE 488.1 compliant standard
command, *TST? is used. This command runs a full self test and returns
one of the following codes:
0 - no tests failed
1 - one or more tests failed
The HP-IB self test consists of the following tests:
ROM checksum
RAM
Battery
Display Assembly
Calibrator
Measurement Assemblies
Communications Assembly (Implicit)
The communications assembly is tested implicitly, in that the command
will not be accepted or return a result unless the HP-IB interface is
functioning correctly.
Refer to “Test Descriptions”, on page 2-54 if you require a description of
each individual test.
When the *TST? command is executed, the screen is cleared. As each test
takes place, the name of the test is listed on the screen. While a test is
running, the message “Testing...” appears next to the name of the test.
As each stage of the test is completed, the message “Testing...” is
replaced by either the message “Passed” or “Failed”.