Omega RD260A Recording Equipment User Manual


 
7 - 8
7.7 Turning Offset Settings Protection On/Off
Configure the recorder so changes can (on) or cannot (off) be made to the offset settings
in the SET mode. And in the list printing, if ON is set, the offset data are printed.
NOTE When protection is turned on, you can set an offset value must be set in the
SET mode. The initial value is ‘OFF
SETTING PROCEDURE:
Enter the SET UP mode by turning ON the power while pressing and holding the [CH
UP] ( ENT) key until the SET UP display appears.
Use the [ ] key to select the display [ ]. Press the [ENT] key.
Using the [ ] key, select either ‘ON’ or ‘OFF’ and then press the [ENT] key. The
display [ ] will appear. The setting is then complete.
Before leaving the SET UP mode, you have to store your new settings. Use the [ ] key
to select the [ ] display, and press the [ENT] key.
Select [ ] to keep your new settings or [ ] to abort the settings, and
press the [ENT] key. After a few seconds, the OPERATION mode will appear.