Sierra SL-800 Drums User Manual


 
35
If you use a USB to RS 232 converter, this could be the problem. Sierra has tested the ASUS and CONRAD (Nr. 982417) and BELKIN
converters which all worked OK, but we cannot confirm that all converters will function properly.
If this does not solve the issues please write down the error. What do you see on the screen? Report what you have tried this far and report
to your Sierra agent or e-mail this to: service@sierrainstruments.com
Setup DUT
Insure that the Cal=Trak flow cell is suited for the range of the DUT. In general, flow instruments are commonly tested at 0-25-50-75-100%
of their rated maximum flow, so the Cal-Trak flow cell selected should be suited for 25% to 100% of the range of the DUT. Currently the
program is not suited to change Cal=Trak flow cells during a certification run. You would have to make 2 runs and merge the data
manually.
Connect a suitable gas supply to the DUT and insure the inlet pressure is correct. Set up the DUT to its operating condition. Please
remember the Cal=Trak only operates under atmospheric pressure conditions. If your DUT requires some back-pressure, place a back
pressure regulator between the DUT and the Cal=Trak.
Insure your system is free of leaks.
Setup Cal=Soft
Go to the tab marked "Operating screen". In this screen you find 4 steps that need to be taken to certify your DUT. The Operating screen
looks like this: