Tektronix vx4380 Drums User Manual


 
Command Descriptions
3–24
VX4380 256-Crosspoint Relay Matrix Module User Manual
Command Response
trig Open the second relay and close the third relay in the
sequence.
trig Open the third relay in the sequence. Place the trigger
subsystem in the idle state.
INITiate
:CONTinuous
N/A
The trigger subsystem is in the idle state.
ABORt
ROUTe:SCAN
TRIGger[:SEQuence][:SOURce]
TRIGger[:SEQuence][:IMMediate]
[:SEQuence][:DELay]
If a scan list has been defined by a [ROUTe:]SCAN command, this command
places the trigger subsystem in the armed state. In this state, a trigger event
enabled by the TRIGger[:SEQuence]:SOURce command opens the currently
closed channel in a scan list and closes the next channel in the list. Triggers
continue to be recognized until an ABORt command is received, which returns
the trigger subsystem to the IDLE state.
In this example, it is assumed that the Option 01 daughter board is installed on a
VX4380 Module, and that the default module name “m1” is in effect.
Command Response
scan (@m1(1:64)) Define a scan list consisting of all relays in section 1 of the
VX4380 Module.
trigger:Sequence:
count 2
Set the number of times to sequence through a scan list (when
an INITiate[:IMMediate] command is received) to 2.
trigger:sequence:
source hold
Disable all trigger sources.
initiate:continuous Place the trigger subsystem in the arm state.
trigger:seq:imm Override the hold placed on triggers. Close channel 1 on the
VX4380.
Command Syntax
Query Syntax
*RST Value
Related Commands
Description
Examples