VXI VT1538A DJ Equipment User Manual


 
Agilent E1538A Enhanced Frequency/Totalize/PWM SCP 19
To measure period at channel 45 as the average of 1000 signal periods:
*RST
TRIGGER:TIMER .2
Alg executes at .2 sec intervals
SENS:FUNC:PER (@145)
ch 45 to measure signal period
SENS:PER:RANGE 1,(@145)
set period range 10µsec - 1sec
SENS:PER:MODE NPER(@145)
set meas interval by n periods
SENS:PER:NPER 1000,(@145)
meas and avg 1000 sig periods
ALG:DEF ’ALG1’,’writecvt(I145,45);’
alg puts period in CVT
INIT
start algorithm execution
do loop
SENS:DATA:CVT? (@45)
get period from CVT
read value from CVT query above
end loop