VXI VT1538A DJ Equipment User Manual


 
Agilent E1538A Enhanced Frequency/Totalize/PWM SCP 55
SENSe:]FUNCtion:TOTalize
[SENSe:]FUNCtion:TOTalize <
ch_list
> sets the SENSe function to
TOTalize for channels in <ch_list>.
See “Totalize Positive or Negative Edge
State Changes” on page 15.
Parameters
Comments The totalize function counts rising edges of digital transitions at
Frequency/Totalize SCP channels. The counter is 24 bits wide and can
count up to 16,777,215.
The SENS:TOT:RESET:MODE command controls which events will
reset the counter.
If the channels specified are not on a Frequency/Totalize SCP, an error
will be generated.
Related Commands: SENS:TOT:RESET:MODE,
INPUT:POLARITY
*RST Condition: SENS:FUNC:COND and INP:POL NORM for all
digital SCP channels.
Send with VXIplug&play Function: hpe14XX_cmd(...)
Usage See “Totalize Positive or Negative Edge State Changes” on page 15.
SENS:FUNC:TOT (@134)
channel 34 is a totalizer
Parameter
Name
Parameter
Type
Range of
Values
Default
Units
ch_list string 100 - 163 none