Reference
80E00 Electrical Sampling Modules User Manual
19
propagation time from the acquisition point to the short in the device under test
and back. See Figure 8.
250 mV
0V
E
i
E
r
Figure 8: St ep generator with a shorted output
Operation Into a 50 Ω Load. Initially, the diode switch is conducting --10 mA.
Since the step-generator output is connected to a 50 Ω load, the resistance to
ground at the acquisition point is 25 Ω (because of the internal 50 Ω impedance).
+250 mV
0V
E
i
E
r
Figure 9: St ep generation with a 50 Ω load
When the diode switch opens (reverse-biased), apparent resistance to ground at
the acquisition point (and at the channel connector) is 25 Ω because the internal
termination resistance is 50 Ω in parallel with the connector impedance of 50 Ω.
The voltage at the acquisition point rises to +250 mV.
The transition propagates to the 50 Ω load and no reflection occurs.
Operation Into an Open. Initially, the diode switch is conducting --10 mA. Since
the step-generator output is open, the resistance to ground at the acquisition point
is 50 Ω (because of the internal 50 Ω impedance).
+250 mV
0V
+500 mV
E
i
E
r
Figure 10: St ep generation with an open circuit