Agilent Technologies 16700 Musical Instrument Amplifier User Manual


 
101
16534A Specifications*
Bandwidth dc to 500 MHz
dc offset accuracy ±(1% of offset + 2% of full scale)
dc voltage measurement accuracy ±(1.5% of full scale + offset accuracy)
Time interval measurement ±[(0.005% of D T) + (2E6 x delay setting) + 100 ps]
accuracy at maximum sampling rate,
on a single scope card, on a single
acquisition
Trigger sensitivity (See notes)
dc to 50 MHz 0.06 full scale
50 MHz to 500 MHz 0.13 full scale
Input resistance 1 M±1% 50 ±1%
* Specifications refer to the input to the BNC connector
Notes:
Specifications apply only within ± 10°C of the temperature at which the most recent calibration was performed.
Specifications apply only after operational accuracy calibration is performed in the frame in which the
oscilloscope module is installed.
Display magnification is used below 56 mV full scale. For sensitivities from 16 mV to 56 mV full scale, full scale is
defined as 56 mV.
Characteristics
General
Maximum sampling rate 2 GSa/s
Number of channels 2 to 8 using the same
time base and trigger.
Up to 10 channels may be installed in a single
16700 frame, or up to 20 in a single system using a
16701 expansion frame.
Waveform record length 32768 points
Oscilloscope Modules Specifications and
Characteristics