State/Timing Modules Specifications and
Characteristics
Agilent Technologies 16710A, 16711A, 16712A
Supplemental Specifications* and Characteristics
Probes (general-purpose lead set)
Input resistance 100 K, ±2%
Parasitic tip capacitance 1.5 pF
Minimum voltage swing 500 mV, peak-to-peak
Threshold accuracy* ±(100 mV + 3% of threshold setting)
Maximum input voltage ±40 V peak
State Analysis
Minimum state clock pulse width 3.5 ns
Time tag resolution [1] 8 ns
Maximum time count between states 34 seconds
Maximum state tag 4.29 x 10
9
states
count between states [1]
Minimum master to master clock time* 16710A, 16711A, 16712A: 10 ns
Minimum master to slave clock time 0.0 ns
Minimum slave to master clock time 4.0 ns
Context store block sizes 16, 32, 64 states
16710A/11A/12A only
Timing Analysis
Sample period accuracy 0.01% of sample period
Channel-to-channel skew 2 ns, typical
Time interval accuracy ± (sample period + channel-to-channel
skew + 0.01% of time interval reading)
Minimum detectable glitch 3.5 ns
* All specifications noted by an asterisk are the performance standards against which the product is tested.
[1] Time or state tags halve the acquisition memory when there are no unassigned pods.
1.5pF
370 ohms
7.4pF
GROUND
100K
ohm
Figure 6.12. Equivalent probe load for the
Agilent 16710A, 16711A and 16712A, general-
purpose lead set.