Agilent Technologies 16700 Musical Instrument Amplifier User Manual


 
95
State/Timing Modules Specifications and
Characteristics
Agilent Technologies 16760A
Supplemental Specifications* and Characteristics
Probes E5378A Single-ended E5379A Differential E5380A Mictor E5382A Single-Ended Flying Leads
Input resistance and Refer to figure 6.14 Refer to figure 6.14 Refer to figure 6.14 Refer to figure 6.15
capacitance
Maximum state data 1.5 Gb/s 1.5 Gb/s 600 Mb/s 1.5 Gb/s
rate supported
Mating connector Agilent part number Agilent part number Amp Mictor 38 [2] None required
1253-3620 [1] 1253-3620 [1]
Minimum voltage swing 250 mV p-p V
in
+
- V
in
-
>= 200 mV p-p 300 mV p-p 250 mV p-p
Input dynamic range -3 Vdc to +5 Vdc -3 Vdc to +5 Vdc -3 Vdc to +5 Vdc -3 Vdc to +5 Vdc
Threshold accuracy +/- (30 mV + 1% of setting)* +/- (30 mV + 1% of setting) [3] +/- (30 mV + 1% of setting) +/- (30 mV + 1% of setting)
Threshold range -3.0 V to +5.0 V -3.0 V to +5.0 V -3.0 V to +5.0 V -3.0 V to +5.0 V
User-supplied threshold -3.0 V to +5.0 V N/A N/A N/A
input range
User-supplied threshold >= 100K ohms N/A N/A N/A
input resistance
Threshold control options User-provided input If operated single-ended Adjustable from user Adjustable from user
Adjustable from user (minus inputs grounded), interface interface
interface the threshold can be adjusted
from the user interface
Maximum nondestructive +/-40 Vdc +/-40 Vdc +/-40 Vdc +/-40 Vdc
input voltage
Maximum input slew rate 5 V/ns 5 V/ns 5 V/ns 5 V/ns
Clock input Differential Differential Single-ended Differential
Number of inputs [4] 34 (32 data and 2 clock/data) 17 (16 data and 1 clock/data) 34 (32 data and 2 clock/data) 17 (16 data and 1 clock/data)
* All specifications noted by an asterisk are the performance standards against which the product is tested.
[1] A support shroud, Agilent part number 16760-02302 (for boards up to 0.062" thick) or 16760-02303 (for boards up to 0.120" thick) is recommended.
A kit of 5 shrouds and 5 connectors is available as Agilent part number 16760-68702 (for boards up to 0.062" thick) or 16760-68703 (for boards up to 0.120" thick).
[2] A kit of 5 Amp Mictor connectors and 5 support shrouds is available, Agilent part number E5346-68701.
A support shroud is available separately, Agilent part number E5346-44701.
[3] If operated single-ended (minus inputs grounded), the threshold can be adjusted from the user interface.
[4] Refer to specifications on specific modes of operation for details on how inputs can be used.
R
1
R
2
C
1
0.7pF
20k
+0.75 V
Model Number C
1
R
1
R
2
E5378A, E5379A 1.5pF 120 30
E5380A 3pF 120 60
Figure 6.14. E5378A, E5379A, E5380A input
equivalent probe load.
121215
30
1pF 0.6pF
20k
+0.75 V
Figure 6.15. E5382A input equivalent probe load, with
5cm damped wire (see users guide for load models