Agilent Technologies 16700 Musical Instrument Amplifier User Manual


 
89
State/Timing Modules Specifications and
Characteristics
Agilent Technologies 16715A, 16716A, 16717A, 16740A, 16741A,
16742A, 16750A, 16751A, 16752A Supplemental Specifications*
and Characteristics
Probes (general-purpose lead set)
Input resistance 100 K, ± 2%
Parasitic tip capacitance 1.5 pF
Minimum voltage swing 500 mV, peak-to-peak
Minimum input overdrive 250 mV
Threshold range -6V to +6V in 10 mV increments
Threshold accuracy* ± (65 mV + 1.5% of settings)
Input dynamic range ± 10V about threshold
Maximum input voltage ± 40V peak
+5V Accessory current 1/3 amp maximum per pod
Channel assignment Each group of 34 channels can be assigned to
Analyzer 1, Analyzer 2 or remain unassigned
2 GHz Timing Zoom (Agilent 16716A, 16717A, 16740A, 16741A, 16742A, 16750A, 16751A, 16752A only)
Timing analysis sample rate 2 GHz/1 GHz/500 MHz/250 MHz
Sample period accuracy ± 50 ps
Channel-to-channel skew < 1.0 ns
Time interval accuracy ± (sample period + channel-to-channel skew + 0.01% of
time interval reading)
Memory depth 16 K
Trigger position Start, center, end, or user defined
Operating Environment
Temperature Agilent 16700 Series frame: 0˚C to 50˚C (+32˚F to 122˚F)
Probe lead sets and cables: 0˚C to 65˚C (+32˚F to 149˚F)
Humidity 80% relative humidity at + 40˚C
Altitude Operating 4600 m (15,000 ft)
Non-operating 15,300 m (50,000 ft)
* All specifications noted by an asterisk are the performance standards against which the product is tested.
1.5pF
370 ohms
7.4pF
GROUND
100K
ohm
Figure 6.13. Equivalent probe load for the
Agilent 16715A, 16716A, 16717A, 16718A,
16719A, 16750A, 16751A, 16752A general-
purpose lead set.