Gardner LCS IV Drums User Manual


 
101
Maintenance
1. Switch off instrument.
2. Remove the power cable from the instrument.
3. Open the cuvette compartment cover and remove cuvette
adapter Z.
4. Using a cross-head screwdriver, loosen the two screws on
the bottom of the cuvette compartment.
5. Lift the cuvette compartment upward and out.
6. Insert the new cuvette compartment so that the guide slots of
the pressure rollers are on the left.
7. Using a cross-head screwdriver, tighten the two screws on
the bottom of the cuvette compartment.
8. Connect the power cable to the instrument.
9. Insert cuvette adapter Z again.