Gardner LCS IV Drums User Manual


 
73
Standard programs
5.7.1.2 Take single wavelength readings (single reading)
1. Insert the zero solution cell into the cell compartment.
2. Press Zero.
Note: The Read key is only active after the zero reading has been
completed.
3. Insert the analysis cell into the cell compartment.
4. Press Read.
5. For data storage, refer to section 5.3.1, page 38.
5.7.2 Multi Wavelength mode – readings with more than one wavelength
In the Multi Wavelength mode, absorbance values can be
measured at up to four wavelengths and the results can be
mathematically processed to obtain sums, differences and
relationships.
Absorbance readings: The light absorbed by the sample is
measured in absorbance units.
% Transmittance measures the percent of the original light
that passes through the sample and reaches the detector.
Turning on the concentration factor allows selection of a
specific multiplier for converting absorbance readings to
concentration. In a graph of concentration versus the
absorbance, the concentration factor is the slope of the line.
Concentration is calculated using a single factor for each
wavelength, which is input by the user.
5.7.2.1 Set the Reading mode at different wavelengths
In the main menu, press the Multi-Wavelength menu option.
Press Options for Parameter Setup.
Table 8 Multi wavelength setup options
Options Description
More For further Options
Save icon
Symbol: Store Data, if Instrument Setup > Data Log Setup > Auto Store: Off is selected.
Symbol: Recall Data, if Instrument Setup > Data Log Setup > Auto Store: On is selected.
% Trans/Abs To switch to % transmittance, concentration or absorbance readings