Commands by Subsystem 3
Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Reference 229
:MEASure:FREQuency
(see page 530)
Command Syntax
:MEASure:FREQuency [<source>]
<source> ::= {CHANnel<n> | FUNCtion | MATH}
<n>::={1|2|3|4}forthefour channel oscilloscope models
<n> ::= {1 | 2} for the two channel oscilloscope models
The :MEASure:FREQuency command installs a screen measurement and
starts a frequency measurement. If the optional source parameter is
specified, the current source is modified.
IF the edge on the screen closest to the trigger reference is rising:
THEN frequency = 1/(time at trailing rising edge - time at leading rising
edge)
ELSE frequency = 1/(time at trailing falling edge - time at leading falling
edge)
Query Syntax
:MEASure:FREQuency? [<source>]
The :MEASure:FREQuency? query measures and outputs the frequency of
the cycle on the screen closest to the trigger reference.
Return Format
<source><NL>
<source> ::= frequency in Hertz in NR3 format
See Also • "Introduction to :MEASure Commands" on page 218
• ":MEASure:SOURce" on page 240
• ":MEASure:PERiod" on page 233
Example Code • "Example Code" on page 240
NOTE
This command is not available if the source is FFT (Fast Fourier Transform).