254 Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Reference
3 Commands by Subsystem
:MEASure:VTOP
(see page 530)
Command Syntax
:MEASure:VTOP [<source>]
<source> ::= {CHANnel<n> | FUNCtion | MATH}
<n>::={1|2|3|4}forthefour channel oscilloscope models
<n> ::= {1 | 2} for the two channel oscilloscope models
The :MEASure:VTOP command installs a screen measurement and starts a
waveform top value measurement.
Query Syntax
:MEASure:VTOP? [<source>]
The :MEASure:VTOP? query returns the vertical value at the top of the
waveform. The top value of the pulse is normally not the same as the
maximum value.
Return Format
<value><NL>
<value> ::= vertical value at the top of the waveform in NR3 format
See Also • "Introduction to :MEASure Commands" on page 218
• ":MEASure:SOURce" on page 240
• ":MEASure:VMAX" on page 249
• ":MEASure:VAMPlitude" on page 246
• ":MEASure:VBASe" on page 248
NOTE
This query is not available if the source is FFT (Fast Fourier Transform).