3 Commands by Subsystem
:SBUS:UART:BASE
(see page 530)
Command Syntax
:SBUS:UART:BASE <base>
<base> ::= {ASCii | BINary | HEX}
The :SBUS:UART:BASE command determines the base to use for the UART
decode display.
Query Syntax
:SBUS:UART:BASE?
The :SBUS:UART:BASE? query returns the current UART decode base
setting.
Return Format
<base><NL>
<base> ::= {ASCii | BINary | HEX}
Errors "- 241, Hardware missing" on page 501
See Also "Introduction to :SBUS Commands" on page 277
":TRIGger:UART Commands" on page 387
NOTE
This command is only valid on 4 (analog) channel oscilloscope models when the
UART/RS-232 triggering and serial decode option (Option 232) has been licensed.