A SERVICE OF

logo

Commands by Subsystem 3
Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Reference 339
:TRIGger:DURation:QUALifier
(see page 530)
Command Syntax
:TRIGger:DURation:QUALifier <qualifier>
<qualifier> ::= {GREaterthan | LESSthan | INRange | OUTRange | TIMeout}
The :TRIGger:DURation:QUALifier command qualifies the trigger duration.
Set the GREaterthan qualifier value with the
:TRIGger:DURation:GREaterthan command.
Set the LESSthan qualifier value with the :TRIGger:DURation:LESSthan
command.
Set the INRange and OUTRange qualifier values with the
:TRIGger:DURation:RANGe command.
Set the TIMeout qualifier value with the :TRIGger:DURation:GREaterthan
command.
Query Syntax
:TRIGger:DURation:QUALifier?
The :TRIGger:DURation:QUALifier? query returns the trigger duration
qualifier.
Return Format
<qualifier><NL>
See Also "Introduction to :TRIGger Commands" on page 313
":TRIGger:DURation:GREaterthan" on page 336
":TRIGger:DURation:LESSthan" on page 337
":TRIGger:DURation:RANGe" on page 340