A SERVICE OF

logo

Commands by Subsystem 3
Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Reference 337
:TRIGger:DURation:LESSthan
(see page 530)
Command Syntax
:TRIGger:DURation:LESSthan <less than time>[<suffix>]
<less than time> ::= maximum trigger duration in seconds
(5 ns - 10 seconds) in NR3 format
<suffix> ::= {s | ms | us | ns | ps}
The :TRIGger:DURation:LESSthan command sets the maximum duration for
the defined pattern when :TRIGger:DURation:QUALifier is set to LESSthan.
Query Syntax
:TRIGger:DURation:LESSthan?
The :TRIGger:DURation:LESSthan? query returns the duration time for the
defined pattern.
Return Format
<less than time><NL>
See Also "Introduction to :TRIGger Commands" on page 313
":TRIGger:DURation:PATTern" on page 338
":TRIGger:DURation:QUALifier" on page 339
":TRIGger:MODE" on page 319