Obsolete and Discontinued Commands 5
Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Reference 489
:MEASure:VDELta
(see page 530)
Query Syntax
:MEASure:VDELta?
The :MEASure:VDELta? query returns the voltage difference between
vertical marker 1 (Y1 cursor) and vertical marker 2 (Y2 cursor). No
measurement is made when the :MEASure:VDELta? query is received by
the oscilloscope. The delta value that is returned is the current value. This
is the same value as the front- panel cursors delta Y value.
VDELta = value at marker 2 - value at marker 1
Return Format
<value><NL>
<value> ::= delta V value in NR1 format
See Also • "Introduction to :MARKer Commands" on page 204
• "Introduction to :MEASure Commands" on page 218
• ":MARKer:Y1Position" on page 211
• ":MARKer:Y2Position" on page 212
• ":MARKer:YDELta" on page 213
• ":MEASure:TDELta" on page 480
• ":MEASure:TSTArt" on page 484
NOTE
The :MEASure:VDELta command is an obsolete command provided for compatibility to
previous oscilloscopes. Use the :MARKer:YDELta command (see page 213) instead.