Procedure to Configure Remote Electronic Track & Hold Board for 220/240 VAC Operation
The board is factory shipped as 110/120 VAC. Perform the following steps to convert the board for use with
220/240 VAC.
1. Remove the zero-ohm resistor jumpers at W1, W3, W4, W6, W7 and W9 on solder side of the remote
electronic track & hold board. Refer to Figure C1.
2 . Install jumpers at W2, W5 and W8 on solder side. Use zero-ohm jumpers or 22 AWG insulated wire.
FIGURE C1 - Remote Electronic Output Board Arrangement