Konica Minolta CM-2500d Drums User Manual


 
E-83
Pass/Fail Judgment for Color Difference
M
emo
If “PASS/FAIL” has been selected as the display mode, “PASS” will be displayed only when all the items have
passed.
If “DIFF/ABS” has been selected as the display mode, the values for the items which have failed will be high-
lighted
If a color difference has not been set, when the target mode is “linked to each data.”, and the target color data
number set at the time of measurement is deleted, “----” appears instead of “PASS/FAIL”.
• If does not appear due to the auto power save function, this may be due the fact that it can take time to
start measurement after C is pressed.
Pass/Fail Judgment Based on Elliptical Tolerances
This function is used to judge whether the measured color difference is within the elliptical tolerances specified
for the target color.
To perform pass/fail judgment based on the specified elliptical tolerances, follow the procedure given below:
Note
To use this function, make sure that elliptical tolerances are set.
“PASS” or “FAIL” is displayed only if “PASS/FAIL” has been selected as the display mode.
[Setting Procedure]
<COND
*
> screen
1
Turn B to select “TARGET”, then press A.
“T*” does not appear if “defined in COND.” mode is selected.
2
Turn B to select “D”, then press A.
D” will change to E.
3
Turn B to select the desired target color no. (T
*
), then press A.
E” will return to “D”.
Keeping B held down to left or right will cause the target color No. to change continuously.