Anderson Manufacturing AV-9000 Recording Equipment User Manual


 
Section 1
1-8
Edition 3
Rev G
To achieve stated results, the following thermocouples must be used with the INPUT TYPE/RANGE set to TC NARROW
4
REF+LIN
+CAL
°C
0.55
1.12
2.98
1.45
1.90
1.87
2.39
TC
TYPE
T
R
S
B
See Note:
RANGE
°C
0/400
-200/0
-250/-220
800/1700
200/800
250/1750
200/1800
1
REF
ACC'Y
°C
0.12
0.23
0.73
0.43
0.58
0.56
0.74
2
LIN.
ACC'Y
°C
0.09
0.19
0.36
0.19
0.25
0.25
0.31
3
FACTORY
CAL
°C
0.34
0.70
1.90
0.83
1.08
1.05
1.34
5
DEVIATION
ACC'Y
°C
0.04
0.02
0.00
0.09
0.06
0.15
0.16
6
RESOL
°C/bit
0.018
0.036
0.111
0.065
0.088
0.086
0.113
ACCURACY TABLES CONT.
4
REF+LIN
+CAL
°C
0.33
0.46
0.32
0.66
0.21
RTD
TYPE
385
DIN
392
USA
392
SAMA
100 ohm
Nickel
See Note:
RANGE
°C
-160/480
-200/-160
-100/450
-200/560
-40/200
1
REF
ACC'Y
°C
0.16
0.14
0.16
0.29
0.09
2
LIN.
ACC'Y
°C
0.03
0.20
0.03
0.13
0.05
3
FACTORY
CAL
°C
0.13
0.12
0.13
0.24
0.07
5
DEVIATION
ACC'Y
°C
0.06
0.00
0.05
0.06
0.02
6
RESOL
°C/bit
0.025
0.022
0.025
0.044
0.013